D6022399
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate far below target
Pass rate is 82.78% (125 of 151 tests passed, 26 failed) against a target of 98%.
5 serial number(s) failed repeatedly
Worst unit 001156519 failed 5 times (last on 2026-08-13 12:07). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001156519 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most frequent failure mode
Error 111 — "DMM: Měření napětí: +3V BAT (TP154/TP120)" occurred 7 times, 26.9% of all 26 failures in the window.
Most-failing measurement step
Measurement step 2.9 — "DMM: Měření napětí: +3V BAT (TP154/TP120)" failed 7 times across the window's 26 failed tests.